T7D UV/VIS Spectrophotometer
Based on the provided document, here are the detailed specifications for the T7D and T7DS UV-Visible Spectrophotometers:
Optical System & Performance
| Parameter | T7D Specification | T7DS Specification |
| Optical System | Dual beam | Dual beam |
| Spectral Bandwidth | 2nm (fixed slit) | 0.5, 1, 2, 5nm (variable slit) |
| Wavelength Range | 190 ~ 1100nm | 190 ~ 1100nm |
| Wavelength Accuracy | $\pm$ 0.3nm (Automatic correction) | $\pm$ 0.3nm (Automatic correction) |
| Wavelength Reproducibility | 0.2nm | 0.2nm |
| Stray Light | < 0.12%T (220nm, NaI; 340nm, $\text{NaNO}_2$) | < 0.12%T (220nm, NaI; 340nm, $\text{NaNO}_2$) |
Photometric Performance
-
Photometric Mode: Transmittance, Absorbance, Energy
-
Photometric Range: -0.3 ~ 3Abs
-
Photometric Accuracy:
-
$\pm$ 0.002Abs (0 ~ 0.5A)
-
$\pm$ 0.004Abs (0.5 ~ 1A)
-
$\pm$ 0.3%T (0 ~ 100%T)
-
-
Photometric Reproducibility:
-
$\pm$ 0.001Abs (0 ~ 0.5A)
-
$\pm$ 0.002Abs (0.5 ~ 1A)
-
$\pm$ 0.15%T (0 ~ 100%T)
-
-
Baseline Flatness: $\pm$ 0.0015Abs (190 ~ 1100nm)
-
Baseline Stability: 0.0008Abs/h (500nm, 0Abs, 2nm Spectral Bandwidth, 2hr warm-up)
-
Photometric Noise: $\pm$ 0.001Abs (500nm, 0Abs, 2nm Spectral Bandwidth)
Operation & Software
-
Working Mode: MPU Mode / PC Mode
-
Software Support: MPU Software Platform / Spec UV software workstation
-
Core Functions: Photometric measurement, quantitative measurement, spectrum scan, kinetics measurement, and DNA/Protein analysis.